ChemSight
The ChemSight™ detector is an infrared (IR), line of sight, sensing system for point of entry, fixed security, fence line and perimeter protection that utilizes non-contact spectral analysis of the atmosphere. Features include; Continuous, monitoring; Long line monitoring; Multi-chemical detection; Digital chemical library updateable via internet: No consumables
Availability: DISCONTINUED
Gabriel Laufer
+1 434 973 1450
1180 Seminole Trail Suite 495
Charlottesville, VA 22901
USA
|
|
www.avirsensors.com
Manufacturer |
| Avir Sensors |
Product Name |
| ChemSight |
Detection Category |
| Chemical; Explosives |
Detection Principle |
| Molecular Spectroscopy; |
Detection Method |
| Mid-infrared Spectroscopy; |
Application |
| Bulk Analysis; Screening; Emergency Response; |
Equipment Type |
| Instrument |
Product Synopsis |
| The ChemSight™ detector is an infrared (IR), line of sight, sensing system for point of entry, fixed security, fence line and perimeter protection that utilizes non-contact spectral analysis of the atmosphere. Features include; Continuous, monitoring; Long line monitoring; Multi-chemical detection; Digital chemical library updateable via internet: No consumables |
Availability |
| DISCONTINUED |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Dimensions |
| Detector: 7 x 7.5 x 22.5 in (178 x 191 x 572 mm); IR lamp: 14 x 14 x 8 in (356 x 356 x 203 mm) |
Weight |
| <30 lbs.; Detector: 18 lbs (8.2 kg); IR lamp: 10 lbs (4.5 kg) |
Power Requirements |
| AC/DC Line Power; 120/240 V AC |
Transportability |
| Person-Portable |
Ruggedness |
| Extreme Temperatures/Humidity |
Operating Conditions |
| –40 to 60°C (–40 to 140°F); 0 to 95% RH (Non-condensing) |
Consumables |
| None |
Calibration Schedule |
| Calibrate after moved. Once calibrated, no need to recalibrate |
Suggested Routine Maintenance |
| System self monitors to determine maintenance, if any. |
Data Analysis Support Equipment |
| Yes |
Communications Interface |
| Can be connected to the internet and controlled from a remote location |
Expected Operational Life |
| Indefinite |
Shelf Life |
| >5 years |
Unit Cost |
| $30,000.00 |
Maintenance Cost |
| Minimum maintenance is required |
Tested For Chemical Agents |
| Hydrogen cyanide, phosgene, arsine |
Tested For ITF-25 TIC/TIMs |
| Ammonia, chloroethane, DIMP, DMMP, ethylene oxide, formaldehyde, hydrogen bromide, hydrogen chloride, hydrogen cyanide, hydrogen peroxide, hydrogen sulfide, phosgene, arsine, sulfur hexafluoride, vinyl chloride |
Other Chemical Targets |
| Butane, carbon dioxide, dibrorane, diesel exhaust, gasoline exhaust, FM-200, IsoClean, isopropynol, methyl salicylate, natural gas, propane, stainless steel cleaner, various alcohols, various solvents |
Adaptability |
| Yes; user can add new targets in the field |
Sample Introduction |
| Vapor/Gas |
Sensitivity/Detection Limits |
| ppm |
Known Interferents/Inhibitors |
| Background rejection capabilities |
Start Up Time (From Cold Start To Sample Ready) |
| Self calibration lasts for 30 minutes. Every time the instrument is moved, calibration must be performed. If not moved, the instrument will start up within a minute. |
Response Time (Sample Application To Output) |
| Continuous (<30 seconds); 1 to 30 seconds |
Total Run Time |
| 3 minutes |
Alarm Capability |
| Visual |
Software Control |
| On-board computer provides gas identification, concentration, time stamp, and other data in XML or flat file format for easy inclusion into industry standard software. |
Training Required |
| <1 week training |
Training Available |
| No |
Manuals Available |
| Operator's manual |
Warranties |
| 2 year |
IATA Shipping Restrictions |
| Yes; Calibration gas cylinder must be shipped per IATA restrictions for compressed gas. |
Data File Type Format |
| iXML, flat file |