Falcon II

ChemImage closed business in April 2023. DISCONTINUED - The FALCON II is a Wide-field Raman Chemical Imaging System optimized for both chemical imaging and spectroscopy. It employs high throughput Liquid Crystal Tunable Filter (LCTF) technology and Electron Multiplication Charge Coupled Device (EMCCD) cameras for Wide-field Hyperspectral Imaging.

Notify me on updates of this product 

Availability: DISCONTINUED

info@chemimage.com
(412) 241-7335
7301 Penn Avenue
Pittsburgh, PA 15208
www.chemimage.com
Manufacturer
ChemImage Sensor Systems (HISTORICAL)
Product Name
Falcon II
Detection Category
Chemical; Explosives;
Detection Principle
Molecular Spectroscopy;
Detection Method
Raman Spectroscopy; Hyperspectral Imaging;
Application
Laboratory Analysis; Bulk Analysis; Screening;
Equipment Type
Instrument
Product Synopsis
ChemImage closed business in April 2023. DISCONTINUED - The FALCON II is a Wide-field Raman Chemical Imaging System optimized for both chemical imaging and spectroscopy. It employs high throughput Liquid Crystal Tunable Filter (LCTF) technology and Electron Multiplication Charge Coupled Device (EMCCD) cameras for Wide-field Hyperspectral Imaging.
Availability
DISCONTINUED
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
36 x 27 x 75 in (92 x 69 x 191 cm)
Weight
>50 lbs.; 650 lbs (295 kg)
Power Requirements
AC/DC Line Power; 110 VAC 60 Hz
Transportability
Benchtop/Fixed
Ruggedness
None; Indoor use only; Not militarized, drop sensitive
Operating Conditions
Indoor use only
Consumables
<$1,000/year
Shelf Life
>5 years
Tested For Chemical Agents
Yes
Tested For ITF-25 TIC/TIMs
Yes
Tested For Explosives
Yes
Other Chemical Targets
Polymers, paint, inks, glass, illicit drugs
Sample Introduction
Solid; Liquid
Sensitivity/Detection Limits
Bulk; mg; g
Start Up Time (From Cold Start To Sample Ready)
< 15 minutes
Response Time (Sample Application To Output)
<15 minutes; 1 to 10 min/real-time imaging
Alarm Capability
No
Software Control
ChemImage Xpert Software
Other Operational Parameters
Optical resolution near diffraction limited to 350 nm Spectral resolution 9 cm-1 FWHM (based on Raleigh criteria) 0.5 cm-1 peak shift discrimination Measured Raman range from 244 cm-1 to 5,464 cm-1 shift from laser line High sensitivity deep cooled back-illuminated CCD –70°C Optional EMCCD (Professional & Standard Series) Maximum optical magnification 400X Excitation source solid-state single frequency
Training Required
<1 week training
Training Available
Yes
Manuals Available
Yes
Warranties
1 year
IATA Shipping Restrictions
No
Please enter contact info.
Email Address: 
Company: 
Name: 
Title: