Model IR-700-CH Combustible Gas Sensor
Detcon Model IR-700-CH is a non-intrusive “Smart” sensor designed to detect and monitor combustible hydrocarbon gases in air over the range of 0-100% LEL using miniature non-dispersive Infrared Optical (NDIR) sensor technology. As compared to catalytic bead sensors, with NDIR there is no risk of sensor poisoning, no risk of high concentration saturation, and no need for O2 to be present. The plug-in, field replaceable detector features over-sized gold-plated connections that help prevent corrosion.
Availability: Commercially Available
Manufacturer |
| Detcon, Inc. |
Product Name |
| Model IR-700-CH Combustible Gas Sensor |
Detection Category |
| Chemical |
Detection Principle |
| Molecular Spectroscopy; |
Detection Method |
| Non-Dispersive Infrared (NDIR); Infrared (IR) Spectroscopy ; |
Application |
| Gas Analysis; |
Equipment Type |
| Sensor |
Product Synopsis |
| Detcon Model IR-700-CH is a non-intrusive “Smart” sensor designed to detect and monitor combustible hydrocarbon gases in air over the range of 0-100% LEL using miniature non-dispersive Infrared Optical (NDIR) sensor technology. As compared to catalytic bead sensors, with NDIR there is no risk of sensor poisoning, no risk of high concentration saturation, and no need for O2 to be present. The plug-in, field replaceable detector features over-sized gold-plated connections that help prevent corrosion. |
Availability |
| Commercially Available |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Dimensions |
| 7 x 2.2 in (178 x 56 mm) |
Weight |
| <5 lbs.; 2 lbs (0.907kg) (sensor only) |
Power Requirements |
| AC/DC Line Power; 11 to 30 VDC, Consumption 68 mA (<1.7 W, normal operation), 85 mA (2 W, maximum) |
Transportability |
| Pocket Size |
Ruggedness |
| Extreme Temperatures/Humidity |
Operating Conditions |
| -40 to 75°C(-40 to 167°F); 0 to 99% RH, non-condensing |
Solvents/Reagents |
| None |
Calibration Schedule |
| Yes |
Suggested Routine Maintenance |
| Minimal |
Available Accessories |
| Remote Alarm Module (Remote Operation and 2 Alarm Relays plus Fault), Hart Integration Module (Hart Communication Protocol version 7.0, HART Registered), Profibus Integration Module (Certified to Profibus DP Standard), Foundation Fieldbus Integration Module (Foundation Fieldbus version 5.01) Wireless Integration Module (Detcon Model RXT-320, SmartWireless® Transceivers) |
Communications Interface |
| 4-20 mA DC, RS-485 MODBUS-RTU and wireless |
Shelf Life |
| >5 years |
Unit Cost |
| $1,345.00 |
Maintenance Cost |
| Minimal |
Other Chemical Targets |
| Combustible Hydrocarbons |
Adaptability |
| No; cannot detect or add new targets |
Sample Introduction |
| Vapor/Gas |
Sensitivity/Detection Limits |
| %; %LEL |
Response Time (Sample Application To Output) |
| Continuous (<30 seconds); T50: < 6 seconds, T90: < 30 seconds |
Alarm Capability |
| Yes |
Software Control |
| Yes |
Training Required |
| <1 week training |
Manuals Available |
| Operator’s Manual |
Warranties |
| Plug in detector 2 years, Transmitter 2 years |