MicroFID II

The MicroFID II Flame Ionization Detector (FID) is a portable instrument designed for HAZMAT detection, confined space entry and soil gas monitoring. MicroFID II provides measurements over a wide range of VOCs, including methane. Its operating range of 0.1 ppm to 50,000 ppm helps define the severity of a hot zone. MicroFID II allows up to 70 hours of continuous use at very low pressure (80 psi) to maximize field operation up-time. Typical applications include chemical incident response, arson accelerant investigations, soil gas monitoring and petroleum products tank entry.

Notify me on updates of this product 

Availability: DISCONTINUED

Rebecca Robertson
DOD/DHS Account Manager
rebecca.robertson@inficon.com
+1 315 434 2580
Two Technology Place
East Syracuse, NY 13057
www.inficon.com
Manufacturer
INFICON
Product Name
MicroFID II
Detection Category
Chemical
Detection Principle
Sensors;
Detection Method
Flame Ionization Detection (FID);
Application
Emergency Response; Gas Analysis;
Equipment Type
Instrument
Product Synopsis
The MicroFID II Flame Ionization Detector (FID) is a portable instrument designed for HAZMAT detection, confined space entry and soil gas monitoring. MicroFID II provides measurements over a wide range of VOCs, including methane. Its operating range of 0.1 ppm to 50,000 ppm helps define the severity of a hot zone. MicroFID II allows up to 70 hours of continuous use at very low pressure (80 psi) to maximize field operation up-time. Typical applications include chemical incident response, arson accelerant investigations, soil gas monitoring and petroleum products tank entry.
Availability
DISCONTINUED
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
13 x 12 x 3 in (331 x 305 x 77 mm)
Weight
<30 lbs.; 11 lbs (5 kg)
Power Requirements
NiMH Battery; 15 hour use/charge
Transportability
Person-Portable
Ruggedness
None
Operating Conditions
0 to 50°C (32 to 122°F); 0 to 100% RH, non-condensing
Consumables
>$1,000/year; Sample Bag, ALTEF, 1L or 3L, Polymer, Septum Hydrogen fuel cylinder, Exhaust frit and housing replacement, Fluoropore filters, Calibration gas, Hydrogen fuel
Solvents/Reagents
Not Specified
Calibration Schedule
With each start up: Zero point calibration and span gas calibration must be done just prior to each day’s analysis
Suggested Routine Maintenance
ProCare annual service $486
Available Accessories
Hydrostik Adapter Kit; Battery Charger; Gas Bag Adapter; Sample probe; Charcoal Filters; Coalescing Filters; Calibration Gas Regulator
Data Analysis Support Equipment
Handheld device (computer, PDA or mobile phone)
Communications Interface
Detection data may be displayed on an optional PDA and uploaded to a PC through its built-in Bluetooth wireless interface.
Expected Operational Life
10+ years
Shelf Life
>5 years; Dependent on calibration gas shelf life, which varies depending on the gas used. Gases must be replaced before expiration.
Unit Cost
$11,495.00
Maintenance Cost
$486 every year
Tested For Chemical Agents
No
Tested For ITF-25 TIC/TIMs
Yes
Tested For Explosives
No
Tested For Narcotics
No
Other Chemical Targets
Volatile organic compounds
Adaptability
Yes; user can add new targets to detector with support
Sample Introduction
Vapor/Gas
Sensitivity/Detection Limits
ppm
Start Up Time (From Cold Start To Sample Ready)
20 minute warm up
Response Time (Sample Application To Output)
Continuous (<30 seconds); 90% of full scale in < 3 seconds
Alarm Capability
Audible and Visual
Software Control
Integrates with LeakDAS® Mobile and GuideWare monitoring software
Training Required
<1 day training
Training Available
Vendor training at customer or vendor facility
Manuals Available
Operator’s Manual
Warranties
1 year warranty, extended warranty available
IATA Shipping Restrictions
Yes; IATA regulations apply for the fuel cylinder and calibration gases