MiniCAMS Series 2001/3001
The MINICAMS® Series 3001 is an automatic, near-real-time continuous air monitoring system using gas chromatography and sample collection with a solid-adsorbent preconcentrator or fixed-volume sample loop. The MINICAMS® Series 3001 is a flexible, expandable system developed specifically for the detecting all CW agents, simulants, and related compounds at levels below the U.S. Surgeon General's 8-hr Time-Weighted-Average (TWA) concentrations and at the proposed Worker Population Limits and Short-Term Exposure Limits for HD, GB, and VX.
Availability: DISCONTINUED
Greg Houston
Sales Manager
+1 205 733 6900 Ext. 204
2148 Pelhem Parkway, Building 400
Pelhem, Alabama 35124
USA
|
|
www.oico.com
| Manufacturer |
| OI Analytical |
| Product Name |
| MiniCAMS Series 2001/3001 |
| Detection Category |
| Chemical |
| Detection Principle |
| Chromatography;Sensors; |
| Detection Method |
| Gas Chromatography; Pulsed Flame Photometric Detector (PFPD);Flame Photometric Detection; |
| Application |
| Gas Analysis; Laboratory Analysis; |
| Equipment Type |
| Instrument |
| Product Synopsis |
| The MINICAMS® Series 3001 is an automatic, near-real-time continuous air monitoring system using gas chromatography and sample collection with a solid-adsorbent preconcentrator or fixed-volume sample loop. The MINICAMS® Series 3001 is a flexible, expandable system developed specifically for the detecting all CW agents, simulants, and related compounds at levels below the U.S. Surgeon General's 8-hr Time-Weighted-Average (TWA) concentrations and at the proposed Worker Population Limits and Short-Term Exposure Limits for HD, GB, and VX. |
| Availability |
| DISCONTINUED |
| Technology Readiness Level (TRL) |
| 9 |
| User Feedback Sources |
| Indifferent/No user feedback |
| Dimensions |
| 12 x 12 x 10 in (30.5 x 30.5 x 25.4 cm) |
| Weight |
| <30 lbs.; 19.8 lbs (9 kg) |
| Power Requirements |
| AC/DC Line Power; 110 V AC 50/60 Hz (600 W) |
| Transportability |
| Mounted |
| Operating Conditions |
| 0 to 40°C (32 to 104°F) |
| Consumables |
| >$1,000/year; Breathing air and high purity hydrogen required for FPD and PFPD, Helium carrier gas |
| Solvents/Reagents |
| None |
| Calibration Schedule |
| Calibrated for specific applications |
| Suggested Routine Maintenance |
| Columns, detectors, and concentrator materials |
| Communications Interface |
| RS-232, RS-485 |
| Shelf Life |
| >5 years |
| Unit Cost |
| $31,125.00 |
| Maintenance Cost |
| ~ $2,500 or less a yr. Pre concentrator tubes are $1,330 or less for a package of 100 (lasts over a year). The V-TO-G pads are $370 for a package of 100. Support gasses Air, Hydrogen, and Nitrogen will be purchased on an as needed basis. |
| Tested For Chemical Agents |
| Yes |
| Tested For ITF-25 TIC/TIMs |
| Yes |
| Other Chemical Targets |
| Yes, petroleum products |
| Sample Introduction |
| Vapor/Gas |
| Sensitivity/Detection Limits |
| ppm |
| Start Up Time (From Cold Start To Sample Ready) |
| 10 minutes |
| Response Time (Sample Application To Output) |
| <1 minute; Near-real-time, depends on GC programming |
| Alarm Capability |
| Yes |
| Software Control |
| Yes |
| Other Operational Parameters |
| Programmable for specific needs |
| Training Required |
| <1 week training |
| Training Available |
| 2 days training for two people at facility in Birmingham, Alabama |
| Manuals Available |
| User manual |
| Warranties |
| 1 year parts and labor |
| IATA Shipping Restrictions |
| Yes; Compressed gas cylinders must be shipped ground only according to IATA restrictions. |