D3S NET
The D3S NET radiation detector is a radioisotope identification device (RIID) for Gamma emitting isotope identification and Neutron detection. The data collected by this radiation detector is not only shown to the user via the mobile phone app (on the phone that comes with the detector) but also sent to a central control screen showing data from every other SIGMA enabled detector on the network. The D3S NET detector unit houses larger detectors that provide identification of isotopes in seconds, alerting the user and your command center of Gamma isotope source, dose and any Neutron detection.
Availability: Commercially Available
Manufacturer |
| Kromek Group plc |
Product Name |
| D3S NET |
Detection Category |
| Radiological/Nuclear; |
Detection Principle |
| Scintillation Detectors; |
Detection Method |
| Scintillation Detectors; |
Application |
| Survey meter;Personal Radiation Detectors; |
Product Model Number |
| D3S NET |
Equipment Type |
| Instrument |
Product Synopsis |
| The D3S NET radiation detector is a radioisotope identification device (RIID) for Gamma emitting isotope identification and Neutron detection. The data collected by this radiation detector is not only shown to the user via the mobile phone app (on the phone that comes with the detector) but also sent to a central control screen showing data from every other SIGMA enabled detector on the network. The D3S NET detector unit houses larger detectors that provide identification of isotopes in seconds, alerting the user and your command center of Gamma isotope source, dose and any Neutron detection. |
Availability |
| Commercially Available |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Supporting Documentation |
| Docs: D3S NET Brochure.pdf; |
Dimensions |
| 5 x 3 x 0.8 in (120 x 80 x 20 mm) |
Weight |
| <1 lb.; 0.33 lbs (150g) |
Power Requirements |
| Lithium Battery; 1100mAh Lithium polymer (12 hours) |
Transportability |
| Pocket Size |
Ruggedness |
| Extreme Temperatures/Humidity; ANSI N42.32 |
Operating Conditions |
| -20 to 50°C (-4 to 122°F) |
Consumables |
| None |
Calibration Schedule |
| SiPM array LED auto-calibration |
Data Analysis Support Equipment |
| Cell phone, PC |
Communications Interface |
| Mini-USB, bluetooth |
Expected Operational Life |
| Indefinite |
Radiological Targets |
| Gamma, neutron |
Number of Library Entries |
| 416 radionuclides |
Adaptability |
| Yes; user can add new targets to detector with support |
Sample Introduction |
| Radioactive Particle |
Sensitivity/Detection Limits |
| CPS; µSv; dose rate: Up to 15µSv/h at 662keV |
Start Up Time (From Cold Start To Sample Ready) |
| Seconds to minutes |
Response Time (Sample Application To Output) |
| Continuous (<30 seconds) |
Total Run Time |
| Continuous |
Alarm Capability |
| Audible, visual, vibration |
Software Control |
| Yes |
Other Operational Parameters |
| Energy range: 30 keV to 3.0 MeV |
Training Required |
| <1 day training |
Training Available |
| No |
Manuals Available |
| User's Manual |
Warranties |
| 1 year |
IATA Shipping Restrictions |
| Yes; Lithium ion battery |
Data File Type Format |
| SPE, CSV |