Direct Reading Dosimeter Checker

It allows for the simultaneous functional testing of up to six Direct-Reading Dosimeters. The Multi-Dosimeter Checker consists of a plastic cylinder with six holes that surround an 8 µCi Cs-137 Exempt Source. Properly charged and zeroed dosimeters are placed in the Dosimeter Checker and exposed for the required period of time, depending on their range. Typically, a six-hour exposure of a properly calibrated dosimeter will yield a reading from 25-35 mR.

Notify me on updates of this product 

Availability: Commercially Available

+1 877 477 6461
417 Main Avenue West
Rolla, North Dakota 58367
USA
www.dosimeter.com
Manufacturer
Arrow-Tech, Inc.
Product Name
Direct Reading Dosimeter Checker
Detection Category
Radiological/Nuclear
Application
Dosimeters;
Product Model Number
DC-DRD6Hole
Equipment Type
Accessory
Product Synopsis
It allows for the simultaneous functional testing of up to six Direct-Reading Dosimeters. The Multi-Dosimeter Checker consists of a plastic cylinder with six holes that surround an 8 µCi Cs-137 Exempt Source. Properly charged and zeroed dosimeters are placed in the Dosimeter Checker and exposed for the required period of time, depending on their range. Typically, a six-hour exposure of a properly calibrated dosimeter will yield a reading from 25-35 mR.
Availability
Commercially Available
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
2.62 X 1.50 in
Weight
<1 lb.; .35 lbs (0.16 kg)
Transportability
Pocket Size
Unit Cost
$240.00
Sample Introduction
Radioactive Particle
Sensitivity/Detection Limits
mR
Response Time (Sample Application To Output)
Continuous (<30 seconds)
Training Required
<1 day training
Please enter contact info.