picoSpin 45/80 MHz
The picoSpin systems come with a 45 or 80 MHz proton (1H) NMR spectrometer. Liquid samples are simply injected into an internal capillary via front-panel fittings. Only 30 microliters of sample fluid is needed to obtain a spectrum. A highly stable, temperature-controlled permanent magnet to ensure easy maintenance-free operation without the need for liquid cryogens.
Availability: Commercially Available
| Manufacturer |
| Thermo Fisher Scientific, Inc. |
| Product Name |
| picoSpin 45/80 MHz |
| Detection Category |
| Chemical |
| Detection Principle |
| Molecular Spectroscopy; |
| Detection Method |
| Nuclear Magnetic Resonance ; |
| Application |
| Laboratory Analysis; |
| Product Model Number |
| 912A0813 / 912A0813 |
| Equipment Type |
| Instrument |
| Product Synopsis |
| The picoSpin systems come with a 45 or 80 MHz proton (1H) NMR spectrometer. Liquid samples are simply injected into an internal capillary via front-panel fittings. Only 30 microliters of sample fluid is needed to obtain a spectrum. A highly stable, temperature-controlled permanent magnet to ensure easy maintenance-free operation without the need for liquid cryogens. |
| Availability |
| Commercially Available |
| Technology Readiness Level (TRL) |
| 9 |
| Market Entry Date |
| 2011 |
| User Feedback Sources |
| MRIGlobal hands-on experience |
| Supporting Documentation |
| Docs: Thermo - picoSpin series II (45&80 MHz).pdf; |
| Dimensions |
| 8 x 5.75 x 11.5 in (20.4 x 14.5 x 29.3 cm) (45 MHz) 14.3 x 10 x 17 in (36.4 x 25.4 x 43.2 cm) (80 MHz) |
| Weight |
| <50 Lbs.; 10.5 Lbs. (4.5 kg) (45 MHz), 43 Lbs. (19.5 kg) (80 MHz) |
| Power Requirements |
| AC/DC Line Power; 115/230 VAC, 50/60 Hz, 150 W, no battery operation |
| Noise Produced |
| Negligible |
| Transportability |
| Person-Portable |
| Ruggedness |
| Indoor use only |
| Operating Conditions |
| Indoor use only |
| Consumables |
| <$1,000/year; Solvents Syringes, Capillary cartridges custom from Thermo |
| Solvents/Reagents |
| Deuterated solvents |
| Calibration Schedule |
| Application dependent |
| Suggested Routine Maintenance |
| User may need to shim magnet and occasionally clean out capillary tube |
| Data Analysis Support Equipment |
| Needs external computer, Metrelab Mnova™ NMR software |
| Communications Interface |
| USB |
| Expected Operational Life |
| Indefinite |
| Shelf Life |
| >5 years |
| Unit Cost |
| $31,245.00 |
| Tested For Chemical Agents |
| Chemical agents that have protons (1H) can be detected |
| Tested For ITF-25 TIC/TIMs |
| TIC/TIM that have protons (1H) can be detected |
| Tested For Explosives |
| Explosives (nitro aromatics, organic peroxides) that have protons (1H) can be detected |
| Tested For Narcotics |
| Narcotics that have protons (1H) can be detected (fentanyl, carfentanil, sufentanil, alfentanil, remifentanil, heroin, ketamine, lorazepam, 3-quinuclidinyl benzilate, U-47700) |
| Other Chemical Targets |
| Chemicals that have protons (1H) can be detected |
| Sample Introduction |
| Liquid; Solid |
| Sensitivity/Detection Limits |
| ppm; ppb |
| Known Interferents/Inhibitors |
| Mixtures containing molecules each with protons will lead to spectra that will need some expert interpretation |
| Start Up Time (From Cold Start To Sample Ready) |
| < 1 hour |
| Response Time (Sample Application To Output) |
| >15 minutes; 1 minute to hours |
| Total Run Time |
| 1 minute to hours |
| Software Control |
| Laptop |
| Other Operational Parameters |
| Larmor Frequency 82 +/- 1 MHz Proton, 45 +/- 1 MHz Proton, Chemical Shift Resolution < 20 ppb FWHM (80 MHz), < 40 ppb FWHM (45 MHz) |
| Training Required |
| <1 week training |
| Training Available |
| Yes |
| Manuals Available |
| Yes |
| Reachback Service |
| No |
| Warranties |
| 5 year |
| ITAR/EAR Export Regulations |
| No |
| IATA Shipping Restrictions |
| No |
| Data File Type Format |
| JCAMP DX |