GT5000 Terra
The GT5000 Terra is capable of measuring up to 50 gases simultaneously, with constant monitoring, online. This analyzer is specifically designed for demanding and active field and on the go use.
Gasmet Technologies newest portable IR Gas Analyzer is the GT5000 (previously the DX4040). It provides a very cost-effective, easy-to-use solution for multi-component gas analysis in ambient air. The GT5000 comes with a 200 gas library for identification and accurate analysis of gases in part-per-million concentrations. Up to 25 gases can be measured simultaneously, and the selection of these gases from the 200 gas list is easy with an intuitive touch screen display. Typical usage areas include industrial hygiene and emergency response situations.
Availability: Commercially Available
| Manufacturer |
| Gasmet Technologies, Inc. |
| Product Name |
| GT5000 Terra |
| Detection Category |
| Chemical |
| Detection Principle |
| Molecular Spectroscopy; |
| Detection Method |
| Mid-infrared Spectroscopy; |
| Application |
| Gas Analysis; |
| Equipment Type |
| Instrument |
| Product Synopsis |
| The GT5000 Terra is capable of measuring up to 50 gases simultaneously, with constant monitoring, online. This analyzer is specifically designed for demanding and active field and on the go use.
Gasmet Technologies newest portable IR Gas Analyzer is the GT5000 (previously the DX4040). It provides a very cost-effective, easy-to-use solution for multi-component gas analysis in ambient air. The GT5000 comes with a 200 gas library for identification and accurate analysis of gases in part-per-million concentrations. Up to 25 gases can be measured simultaneously, and the selection of these gases from the 200 gas list is easy with an intuitive touch screen display. Typical usage areas include industrial hygiene and emergency response situations.
|
| Availability |
| Commercially Available |
| Technology Readiness Level (TRL) |
| 9 |
| Market Entry Date |
| October, 2018 |
| User Feedback Sources |
| Indifferent/No user feedback |
| Dimensions |
| 17.7 x 11.3 x 6.5 in (45 x 28.7 x 16.6 cm) |
| Weight |
| <30 lbs.; 20.7 lbs. (9.4 kg) with battery, 17.6 lbs. (8.0 kg) without battery |
| Power Requirements |
| Lithium Battery; AC/DC Line Power; Li-ion battery (3 hours) or 115/230 VAC |
| Noise Produced |
| Minimal |
| Transportability |
| Person-Portable |
| Ruggedness |
| Waterproof/Resistant; Dust; IP54 |
| Operating Conditions |
| Short term temperature: 0 to 40°C (32 to 150°F) Long term temperature: 5 to 30°C (41 to 86°F); 90%RH, non-condensing |
| Consumables |
| <$1,000/year; Gas filters (2µm PTFE filter) |
| Solvents/Reagents |
| None |
| Calibration Schedule |
| Zero point calibration performed using nitrogen (4.0 or higher N2 recommended) immediately before sampling and no more than once every 24 hours |
| Available Accessories |
| Gasmet sampling probe with 2 µm PTFE filter |
| Data Analysis Support Equipment |
| Sample Cell Portable Sampling System Calibrator Pressure measurement: Inside sample cell Analog Signals (ext PC): ADAM 5000/TCP Sample cell gaskets: Teflon® coated Viton® or Kalrez® -Trolley: wheeled cart for the analyzer and laptop computer |
| Communications Interface |
| Bluetooth, RS232 Analyzer is connected to an external computer via RS- 232C cable. The external computer controls the instrument. The system operating software is CALCMET for Windows XP. |
| Expected Operational Life |
| Indefinite |
| Shelf Life |
| >5 years |
| Unit Cost |
| $60,000.00 |
| Tested For Chemical Agents |
| Yes |
| Tested For ITF-25 TIC/TIMs |
| Yes |
| Other Chemical Targets |
| Yes |
| Number of Library Entries |
| 300+ |
| Adaptability |
| Yes; user can add new targets to detector with support |
| Sample Introduction |
| Vapor/Gas; |
| Sensitivity/Detection Limits |
| % |
| False Positive Rates |
| <5 % |
| False Negative Rates |
| <5 % |
| Response Time (Sample Application To Output) |
| <5 minutes; <120 seconds |
| Total Run Time |
| Continuous monitor |
| Alarm Capability |
| Yes |
| Software Control |
| On-board, PC |
| Other Operational Parameters |
| 8 cm-1 resolution, 2LPM flow, 6 mm quick-connect, Wave number range: 900 - 4 200 cm-1, |
| Training Required |
| <1 day training |
| Training Available |
| Vendor training at customer or vendor facility |
| Manuals Available |
| Yes |
| Warranties |
| Yes |
| ITAR/EAR Export Regulations |
| No |
| IATA Shipping Restrictions |
| Yes; Lithium-ion battery |